beautypg.com

Error messages on the display, F01 to f12 – Retsch PM 200 User Manual

Page 34

background image

Retsch GmbH

34

Error messages on the display

F01 to F12

F02

The display shows

F01

Lock indicator

Close cover

The error appears when the cover

has not been fully closed, or if the

trannion magnets have not been

recognised.

Close the cover

The display shows

The display shows

F03

F04

Problem in the

Close or open cover,

safety circuit of

should problem remain

the interlock

then interlock faulty

Service required!

Service required!

The error appears when the

interlock is defective.

Turn the device off at the

main switch, there is a

safety problem.

The error appears when the

interlock is defective.

Turn the device off at the

main switch, there is a

safety problem.

F05

F06

The display shows

The display shows

F07

F08

Faulty motor

Please select

control

grinding time

Service required!

The display shows

The display shows

F09

F10

Cooling fan is off

Sun wheel does

not rotate

Service required!

Service required!

F12

The display shows

F11

Sun wheel rotate

to slow

Service required!

This manual is related to the following products: