Error messages on the display, F01 to f12 – Retsch PM 200 User Manual
Page 34
Retsch GmbH
34
Error messages on the display
F01 to F12
F02
The display shows
F01
Lock indicator
Close cover
The error appears when the cover
has not been fully closed, or if the
trannion magnets have not been
recognised.
Close the cover
The display shows
The display shows
F03
F04
Problem in the
Close or open cover,
safety circuit of
should problem remain
the interlock
then interlock faulty
Service required!
Service required!
The error appears when the
interlock is defective.
Turn the device off at the
main switch, there is a
safety problem.
The error appears when the
interlock is defective.
Turn the device off at the
main switch, there is a
safety problem.
F05
F06
The display shows
The display shows
F07
F08
Faulty motor
Please select
control
grinding time
Service required!
The display shows
The display shows
F09
F10
Cooling fan is off
Sun wheel does
not rotate
Service required!
Service required!
F12
The display shows
F11
Sun wheel rotate
to slow
Service required!