Draining the reservoir to an external location, Draining the reservoir to an external location 44, Draining the reservoir – MTS Model 286-20 Confining Pressure Intensifier User Manual
Page 44: To an external location, Procedure, Draining the reservoir to an external container, 20 pressure intensifier, Operating procedures operation, 2 - b

286.20 Pressure Intensifier
44
Operating Procedures
Operation
Draining the Reservoir to an External Location
Procedure:
1. Connect a hose between the connector on the lower side of the Fluid
Direction valve (rear side of control panel) and the external fluid container.
2. Set the Fluid Direction valve to position B. Confining fluid in the fluid
reservoir will flow into the container by means of gravity.
Triaxial
Cell
Open
Close
Confining Fluid Input
Open
Close
Confining Fluid Output
Open
Close
Reservoir Fill
Fluid
Reservoir
Open
Close
Intensifier Fill
Pressure
Intensifier
Low Pressure
Relief
Pressure
Transducer
Low pressure line
High pressure line
286.20
Confining
Pressure
Intensifier
r
Fluid Filter
Off
Fluid Transfer Pump Control
Air supply in
100 psi max.
Air Driven
Hydraulic
Pump
Fluid Direction
A
B
C
C
A
B
Empty
Reservoir
From fluid
supply or to
empty fluid
Fill system from
fluid supply
using pump
Fill system
from reservoir
using pump
Confining
Pressure
A
B
A
B
Air Direction
Air supply into empty
traxial cell.
Bleed air into reservoir
when filling triaxial cell.
psi
bar
High Pressure
Relief
Air Pressure
Relief
Draining the Reservoir To An External Container
2 - B