286.20 Pressure Intensifier
20
Control Panel
Operation
Control Panel
The following figure shows the operational controls, superimposed on the front
panel hydraulic schematic. The following table describes items 1 through 8,
shown in the figure.
The table lists the operational effects of each individual control. However, note
that performing the various operations listed in the table usually involves the
manipulation of two or more of the controls.
I
TEM
C
ONTROL
F
UNCTION
1
Confining Fluid
Output
Adjustable hand-valve. Turning the valve counterclockwise toward Open
allows air to be vented from the triaxial cell to the reservoir when filling the
cell with fluid. The valve must be closed before pressurizing the triaxial cell.
2
Confining Pressure
gage
Analog gage. Indicates pressure inside the Triaxial Cell, scaled in pounds
per square inch (psi).
Triaxial
Cell
Open
Close
Confining Fluid Input
Open
Close
Confining Fluid Output
Open
Close
Reservoir Fill
Fluid
Reservoir
Open
Close
Intensifier Fill
Pressure
Intensifier
Low Pressure
Relief
Pressure
Transducer
Low pressure line
High pressure line
286.20
Confining
Pressure
Intensifier
r
Fluid Filter
Off
Fluid Transfer Pump Control
Air supply in
100 psi max.
Air Driven
Hydraulic
Pump
Fluid Direction
A
B
C
C
A
B
Empty
Reservoir
From fluid
supply or to
empty fluid
Fill system from
fluid supply
using pump
Fill system
from reservoir
using pump
Confining
Pressure
A
B
A
B
Air Direction
Air supply into empty
traxial cell.
Bleed air into reservoir
when filling triaxial cell.
psi
bar
High Pressure
Relief
Air Pressure
Relief
1
2
4
3
5
6
8
7