Labnet Prism Air-Cooled Microcentrifuge (C2500) User Manual

Page 8

background image

Adapter for 0.5ml tubes (pk./6)
Order

no.

C-1205

Tube measurement

8 x 30mm

Max.

speed

15,000rpm

Centrifuging radius

7.53cm

RCF (g-value)

18,942 x g

Adapter for 0.4ml tubes (pk./6)
Order

no.

C-1206

Tube measurement

6 x 47mm

Max.

speed

15,000rpm

Centrifuging radius

8.4cm

RCF (g-value)

21,200 x g

Adapter for 0.2ml tubes (pk./6)
Order

no.

C-1222

Tube measurement

6 x 21mm

Max.

speed

15,000rpm

Centrifuging radius

7.03cm

RCF (g-value)

17,684 x g

3.2 Rotor maintenance

The rotor should be cleaned thoroughly after each use. Thorough cleaning must be
performed when spinning samples containing phenol or phenol chloroform.
Periodically inspect the rotor for dents, dings, scratches, discoloration and cracks. If any
damage to the rotor is found, discontinue use of the rotor immediately and replace.

3.3 Removing and installing the angle rotor

The Prism comes complete with a standard 24 place rotor installed. To remove the rotor
for cleaning, remove the rotor securing screw from the motor shaft by turning the screw
counterclockwise, using the rotor wrench. Lift the rotor directly upward in a straight
vertical motion.