5.3 Cleaning the rotor
The rotor should be cleaned after each use. When spinning samples containing phenol or
phenol chloroform, the rotor should be cleaned immediately after use.
5.4 Disinfection
Should a spill of infectious materials occur within the rotor or chamber, the unit should
be disinfected. This should be performed by qualified personnel with proper protective
equipment.
5.5 Replacing fuses
Check the fuse when it is recommended in the Troubleshooting Guide located in this
manual. The fuse holder is located in the power inlet on the rear of the unit. Disconnect
the power cord from the power inlet. Open the fuse holder drawer by inserting a small
screwdriver under the tab and prying it open. Remove the innermost (operative) fuse
from its retaining tabs and replace the fuse if necessary. A spare fuse is located in the
outermost chamber of the fuse drawer. Replace only with a fuse of exactly the same
value as the original. (Fuse type may be found in the
Technical data section of this manual
6. Troubleshooting Guide
Please refer to this guide before calling for service.
Centrifuge will not start
Possible reason:
No power supply
Solution:
Check that power is being supplied to
the
outlet
Check that the power cord is plugged
into both the wall outlet and the back of
the
centrifuge
Check that power cord is not damaged
Possible reason:
Blown fuse
Solution:
Check fuse and replace if necessary
Lid lock will not release
Possible reason:
Defective lid lock
Solution:
Open manually and have unit serviced
Possible reason:
No power from PC board
Solution:
Call for service
Possible reason:
Lid lock is jammed
Solution:
Call for service
Possible reason:
Centrifuge is not receiving power
Solution:
See “Centrifuge will not start”