Centrifuge cannot be started, although power is on

Possible reason:

Lid not closed correctly

Solution

Close lid correctly

Possible reason:

No speed or time has been selected

Solution:

Set speed and/or time

bAL: Error Message

Indicates imbalance

Possible reason:

Tubes not inserted symmetrically in rotor

holes

Solution:

Load tubes symmetrically

(see Section 3.4 on Loading the rotor)

Possible reason:

Sample liquid in tubes not balanced

Solution:

Make sure the same volume of liquid is

in

each

tube

Possible reason:

Defective or improperly adjusted

balance

sensor

Solution: Call

service

Lid: (Error message)

Lid not closed

Possible reason:

Lid not closed completely

Solution: Close

lid

Possible reason:

Lid lock or sensor defective

Solution: Call

service

Other Error Messages

Er on Display

Solution:

Press time or speed knob to clear error

Call

service