Labnet Prism Air-Cooled Microcentrifuge (C2500) User Manual
Page 14

Centrifuge cannot be started, although power is on
Possible reason:
Lid not closed correctly
Solution
Close lid correctly
Possible reason:
No speed or time has been selected
Solution:
Set speed and/or time
bAL: Error Message
Indicates imbalance
Possible reason:
Tubes not inserted symmetrically in rotor
holes
Solution:
Load tubes symmetrically
(see Section 3.4 on Loading the rotor)
Possible reason:
Sample liquid in tubes not balanced
Solution:
Make sure the same volume of liquid is
in
each
tube
Possible reason:
Defective or improperly adjusted
balance
sensor
Solution: Call
service
Lid: (Error message)
Lid not closed
Possible reason:
Lid not closed completely
Solution: Close
lid
Possible reason:
Lid lock or sensor defective
Solution: Call
service
Other Error Messages
Er on Display
Solution:
Press time or speed knob to clear error
Call
service