The centrifuge should be installed on a rigid, even surface such as a stable laboratory
bench, countertop, etc. To guarantee sufficient ventilation, ensure that the centrifuge has
at least 15cm (6 inches) of free space on all sides, including the rear.
The centrifuge should not be located in areas subject to excessive heat such as in direct
sunlight or near radiators or the exhaust of a compressor, as a buildup of heat may occur
within the chamber.

2.3 Installing the Centrifuge

Before operating the centrifuge, check that the power source (electrical outlet on the wall)
corresponds to that on the manufacturer’s rating label, then connect the power cord to the
centrifuge and the power source.

3. Installation of rotors and rotor maintenance

3.1 Rotors and accessories

The following accessories are included or available for the Prism:

Angle rotor for 24 x 1.5ml tubes

(Included)

Order

no.

C2400-2

Tube measurement

1.5ml / 2.0ml (10 x 40mm)

Max.

speed

15,000rpm

Centrifuging radius

8.4cm

Max. RCF (g-value)

21,200 x g

Rotor

Wrench

(Included)

Order

no.

C0240-RT

Rotor

Lid

(Included)

Order

no.

C0240-RC

Rotor Securing Screw

(Included)

Order

no.

C0240-RSS


Strip Spin Adapter
Order

no.

C2400-SS

Tubes accepted

0.2ml tubes or 8 x 0.2ml strips

Max

Speed

15,000

rpm

Range Centrifuge radius

4.32cm to 5.33cm

Range RCF (g value)

10,874 x g to 13,405 x g