JUMO 902425 HEATtemp Screw-In RTD Temperature Probe for Heat Meters with Connecting Cable for Direct Mounting, Type DS/DL Data Sheet User Manual
Technical data

Data sheet 902425
Page 1/10
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
E-mail:
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
E-mail:
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax:
315-437-5860
E-mail:
Internet: www.jumousa.com
Screw-in RTD Temperature Probe
for Heat Meters with Connecting Cable
for Direct Mounting (Type DS/DL)
For temperatures between 0 and 180 °C
Approval for heat meters as replaceable temperature probes;
MID and domestic approvals
Fulfills the requirements of DIN EN 1434, AGFW FW202, and FW211
Approval for cold and combined cold/heat meters
Paired and calibrated according to the German calibration regulations
Production certified according to module D of MID (CE and metrology identification
Heat meter RTD temperature probes are installed to measure temperatures in closed pipeline
systems. For optimum thermic coupling to the measured medium in the heating system, direct
mounting without additional thermowell is recommended.
The temperature probes are approved according to the EU directive 2004/22/EC (MID) as well
as according to the German calibration regulations (domestic approval).
For pipelines with a nominal width smaller than 50 mm, type DS temperature probes are used.
Installation in ball valves has been shown to be the optimum installation location, see data sheet
902440. In addition, emptying the system when installing and replacing the temperature probe
at the end of the calibration period is no longer necessary.
The temperature probes are available in paired and calibrated versions. The officially recognized
testing agency for heat KF2 is available for this. Moreover, the production is certified according
to module D of MID, and the temperature probes are also available with the CE and metrology
identification marking.
Technical data
Connection
Ferrules
Connecting cable
PVC, PUR, TPE, silicone; unshielded or shielded
For max. connecting cable length see table on the next page
Process connection
Type DS: screw connection M10 × 1, brass according to DIN EN 1434
Type DL: thread G 1/4, G 1/2 stainless steel
Protection tube
Type DS: stainless steel Ø 5.4 mm offset by Ш 3.3/Ш 3.6 mm
Type DL: stainless steel Ø 8 mm offset by Ø 6 mm
Measuring insert
Platinum temperature sensor according to DIN EN 60751
Nominal value: Pt100, Pt500, Pt1000; class B
Connection: two-wire/four-wire circuit
Temperature
0 to 180 °C
Temperature difference
3 to 180 K
Minimum immersion depth
Type DS: 15 mm
Type DL: 30 mm
Insertion length
Type DS: 25 to 60 mm
Type DL: 60 to 280 mm
Maximum pressure
PN 25
Flow velocity
2 m/s (water)
Response behavior
Type DS: t
0.5
= 2 s
Type DL: t
0.5
= 6 s
Thermoelectric voltage
≤
5 µV
Environmental conditions
Climatic 0 to 70 °C
Protection type IP65
Electromagnetic E1
Mechanical M3
2014-04-02/00311505