1 Equipment and intended application

RTD temperature probes from JUMO are used as operating equipment for temperature measurement in
liquid or gaseous media, as well as dusts, with the protection types: intrinsic safety “i”, flameproof enclosure
“d” or increased safety “e”. The thermometers consist of a protection fitting with various process connection
options, a terminal head or a connecting cable, and an interchangeable measuring insert (depending on the
type). All fittings in contact with the process media undergo a leakage test. The Pt100 temperature sensors
mounted in the fittings are to EN 60751, in tolerance classes A or B in 2-, 3-, or 4-wire circuit (see Section 8).
It is also possible to use these sensors with a higher reference value (Pt500; Pt1000; Pt2000; Pt 5000). NTC
sensors, such as KTY, or other PTC sensors can also be used. Versions with two or three measurement
circuits are possible. A transmitter can be built into the terminal head, for measurement transmission with a
standard signal (e.g. 4 to 20 mA).

The thermometers fulfill the requirements for Explosion Group II in Categories 1 G and 1 D, as well as 2 G
and 2 D. They are thus suitable for use in areas with an explosion hazard: Zone 1 and 2 with gas (Gas) and
Zone 21 and 22 with dust (Dust). Depending on the circumstances, it may be permissible for the probe stem
to protrude into Zone 0 or 20 (zone separation).

The thermometer parameters that are specific to a given type can be found in the corresponding
technical data sheet/detail drawing (see attachment) and/or on the label affixed to these operating
instructions.


Depending on the measurement task and the application requirements, the RTD temperature probes are
available with different terminal heads, a variety of process connections, matching pockets, with or without
an interchangeable insert, or with a fitted connecting cable.

RTD temperature probes with type Ex “i” protection are certified for connection to intrinsically safe circuits in
Category ib (for use in Zones 1 and 2, with an isolator in Zone 0) and in Category ia (where the probe stem
is in Zones 0, 1 and 2).

RTD temperature probes in a flameproof enclosure (terminal head and cable gland) as well as with
“Increased safety” are also equipped with intrinsically safe measuring inserts for connection to intrinsically
safe circuits.
When connecting up to circuits that are not intrinsically safe, the user must ensure that the power applied is
limited in such a way that the maximum rise in surface temperature is no greater than that specified for the
temperature class minus the safety reserve.
See also Section 5 “Technical data” of these operating instructions.

2 Marking

The type of the RTD temperature probe used is marked on the terminal head or the permanently attached
nameplate. For every RTD temperature probe that carries an individual type designation (e. g.
90.2800.9101), there is a corresponding probe-specific drawing or data sheet. The appropriate marking
ensures that all sensors are unambiguously identifiable and traceable. The probe-specific values can be
found in the drawing, the technical data sheet and/or the label that is affixed to these operating instructions.

3 Safety notes

The technical data that are relevant to the use of the device in an area with an explosion hazard (Ex area)
are presented in the detail drawing/corresponding data sheet and/or the label that is affixed to these
operating instructions.

Only operate RTD temperature probes according to their intended use and in a clean and undamaged state!

It is not permissible to make any alterations to the RTD temperature probes. In such a case, proper fault-free
operation is no longer assured. The guarantee is also invalidated by any alterations. When exchanging
measuring inserts, use only original JUMO replacement parts of the same type.

The national and international safety and accident prevention regulations must be followed for installing and
working on or with the RTD temperature probes. Furthermore, the operator of the plant or installation is
responsible for the observance of legal requirements. If lead extensions are used, the specified (length-
dependent) limits for capacitance and inductance must be observed.