JULABO ME-26 Heating Circulators User Manual
Page 59
ME
59
Command
Parameter
Response of circulator
in_mode_01 none
Selected
setpoint:
0 = SETPNT 1
1 = SETPNT 2
2 = SETPNT 3
3 = Last setpoint setting was carried out
through an external programmer
in_mode_04 none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05 none
Circulator in Stop/Start condition:
0 = Stop
1 = Start
in_mode_08 none
Adjusted control dynamics
0 = aperiodic
1 = standard
12.4. Status
messages
Status messages
Description
00 MANUAL STOP
Circulator in „OFF“ state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in „r OFF“ state.
03 REMOTE START
Circulator in remote control mode.
12.5. Error
messages
Error messages
Description
-01 LOW LEVEL ALARM
Low liquid level alarm.
-02 REFRIGERATOR ALARM
Control cable of the refrigerated circulator or MVS solenoid
valve controller short-circuited or interrupted.
-03 EXCESS TEMPERATURE
WARNING
High temperature warning.
-04 LOW TEMPERATURE
WARNING
Low temperature warning.
-05 WORKING SENSOR ALARM
Working temperature sensor short-circuited or interrupted.