Preparations

16

5. Preparations

5.1. Installation

 Place the circulator in an upright position.
 Place the unit on an even surface on a pad made of non-

flammable material

5.2.

Bath fluids

Caution:

Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of

65 °C is used, only supervised operation is possible.

Water: The quality of water depends on local conditions.
 Due to the high concentration of lime, hard water is not suitable for temperature

control because it leads to calcification in the bath.

 Ferrous water can cause corrosion - even on stainless steel.
 Chloric water can cause pitting corrosion.
 Distilled and deionized water is unsuitable. Their special properties cause

corrosion in the bath, even in stainless steel.

Recommended bath fluids:

Bath fluid

Temperature range

soft/decalcified water

5 °C to 80 °C

JULABO bath fluids

JULABO
Description

Thermal
G

Thermal
M

Thermal
HS

Order Number

10 liters

8 940 124

8 940 100

8 940 102

5 liters

8 940 125

8 940 101

8 940 103

Temperature range

°C

-30 ... 80

40 ... 170

20 ...250

Flash point

°C

--

284

270

Fire point

°C

--

306

>360

Color

light yellow

clear