9.5.2. Dynamic

internal

°C

t

APER

NORM

Setpoint

temp. stability

°C

t

APER

NORM

temperature ramp

temp. stability

This parameter affects the temperature sequence only in
case of internal control.

Factory setting:

APER (aperiodic)

Possible parameters:

NORM

Allows for reaching the setpoint faster – with

setpoint change or ramp function – but
overshooting of up to 5 % is possible.

APER

Ramp function: the increase of temperature
occurs temporally offset and achieves the
target temperature without overshooting.
Setpoint change: The temperature increases at
the same rate, the target temperature is
achieved without overshooting.

With both settings constant temperature is achieved
after approximately the same time.

9.5.3.

Control parameters – XPU-, XP-, TN-, TV- EXTERNAL

Setting range: 0.1 ...99.9

Setting range: 3 ...9999

Setting range: 0 ... 999

In most cases the control parameters preset in the factors are adequate
for achieving an optimum temperature sequence.
The control parameters allow adjustment to special control processes.

Setting range: 0.1 ... 99.9

Proportional range >Xpu<

The proportional range Xpu of the cascaded controller is only needed for
external control.