beautypg.com

JULABO ME-26 Heating Circulators User Manual

Page 18

background image

Preparations

18

5.1.1. Tubing

Recommended tubing:

Order No.

Length

Temperature range

8 930 008

1 m

CR

®

tubing

8 mm inner dia.

-20 °C to 120 °C

8 930 010

1 m

CR

®

tubing

10 mm inner dia.

-20 °C to 120 °C

8 930 108

1 m

Viton tubing 8 mm inner dia.

-50 °C to 200 °C

8 930 110

1 m

Viton tubing 10 mm inner dia.

-50 °C to 200 °C

8 930 410

1 m

Insulation for tubing 8 mm or
10

mm

inner

dia.

-50 °C to 100 °C

8 970 480

2 tubing clamps. size 1, tubing 8 mm inner dia.

8 970 481

2 tubing clamps. size 2, tubing 10 or 12 mm inner dia.

8 930 209

0.5 m

8 930 210

1.0 m

8 930 211

1.5 m

8 930 214

3.0 m

Metal tubing, triple insulated,
M16x1 *

-100 °C to +350 °C

8 930 220

0.5 m

8 930 221

1.0 m

8 930 222

1.5 m

8 930 223

3.0 m

Metal tubing, insulated, M16x1 *

-50 °C to +200 °C

*) Adapter for metal tubing M10x1 on M16x1

Order No. 8 970 444


Warning:

Tubing:

At high working temperatures the tubing used for temperature application and cooling
water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a short time.
This may result in:
 Burning of skin

 Difficulties in breathing due to hot atmosphere

Safety recommendations

 Employ suitable connecting tubing.

 Make sure that the tubing is securely attached.
 Avoid sharp bends in the tubing, and maintain a sufficient distance from

surrounding walls.

 Regularly check the tubing for material defects (e.g. for cracks).

 Preventive maintenance: Replace the tubing from time to time.


This manual is related to the following products: