JULABO ME-26 Heating Circulators User Manual
Page 19
ME
19
5.4.
Filling / draining
Notice:
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit on for
a short moment, for example).
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit, or before moving the unit.
Store and dispose the used bath fluid according to the laws for environmental
protection.
Filling
Take care that no liquid enters the interior of the circulator.
Cooking in the Sous-Vide method
The recommended minimum filling level with water as bath fluid is
50 mm below the tank rim.
Recommended maximum filling level with water as bath fluid:
30 mm below the tank rim
Recommended maximum filling level with bath oils:
40 mm below the tank rim
After filling, immerse the samples in the bath or place the lid on
the bath, in case the opening is not to be used.
The circulator provides an early warning system for low level
(description – please refer to page 28) that may be triggered when
changing samples in the bath.
Draining
Turn the unit off with the mains switch.
Place the heating circulator near the rim of the table. Use a
suitable vessel as recipient for the bath fluid.
Unscrew the drain plug on the front of the bath tank to drain the
bath.