9.9. Analog

inputs/outputs

If the electronic module is not fitted, the item >inputs/outputs< will be
missing in the menu.

ALARM

REG+
E-PROG

STAND-BY

In order to use the analog inputs and outputs, the circulator must be
equipped with the electronic module, which is available as option.
Order No. 8 900 100 Electronic module

This submenu enables the input and output values to be set for the
programmer input and the temperature recorder outputs of the REG+E-
PROG socket.

The >STAND-BY< input and the >ALARM< output are configurable.

Menu item >SV-OUT< is not used.

Channel 1

voltage output for recorder (V)

Channel 2

voltage output for recorder (V))

Channel 3

current output for recorder

(mA)

EPROG

external programmer input

External standby

Standby input

(for external switch-off)

Alarm output

Output for external alarm signal


NOTE:
Channel 2 is pre-configured for the connection of an HST Booster Heater.

Press the

key if a parameter is to be retained. (prior to OK)

Level 2

Parameter-Level

Menu item >SV-OUT< is not used.

Level 2

Level 3

Parameter-level

Define the output value for channel 1

Possible parameters:
ActInt, ActExt, Power, Setpoint

The parameter flashes, switch by pressing