9.6. Serial
Interface
Factory settings:
4800
Baud
even
hardware
handshake
For communication between the circulator and a PC or a superordinated
process system, the interface parameters of both units must be identical.
The adjustment is usually carried out one time only in the >Interface serial<
menu.
RS232
RS485
Level 2
Parameter-level
Press the
key if a parameter is to be
retained. (prior to OK)
The parameter flashes, switch by pressing
and
Possible parameters: RS232 / RS485
Address RS485: 0 to 127
The parameter flashes, switch by pressing
and
Possible parameters:
4800 / 9600 / 19200 / 38400
The parameter flashes, switch by pressing
and
Possible parameters: no/even/odd
even: Data bits = 7; Stop bits = 1
odd:
Data bits = 7; Stop bits = 1
no:
Data bits = 8; Stop bits = 1
The parameter flashes, switch by pressing
and
Possible parameters: hardw / softw
Xon/Xoff-protocol (Software handshake)
Protocol RTS/CTS (Hardware handshake)