Technical specifications – JULABO FK31-SL Calibration Baths User Manual
Page 10
Operator responsibility – Safety recommendations
10
2.4. Technical
specifications
Calibration Baths
FK30-SL
FK31-SL
Working temperature range
°C
-30 ... 200
-30 ... 200
Temperature stability
(1
°C
±0.005
±0.005
Temperature uniformity
°C
<±0.01
<±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20
0.46 0.34 0.15
+20 0 -20
0.46 0.34 0.15
Refrigerant
R134a
R134a
Overall dimensions (WxDxH)
cm
32x45x79
32x45x91
Bath opening (WxL)
cm
12
12
Bath depth
cm
17
31
Filling volume
Liter
14
24
Weight
kg
48
51
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current draw (at 230 V)
Current draw (at 230 V)
A
A
A
14
10 (Circulator)
4 (Cooling machine)
14
10 (Circulator)
4 (Cooling machine)
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current draw (at 115 V)
A
A
10 (Circulator)
6 (Cooling
machine)
10 (Circulator)
6 (Cooling
machine)
(1
±0.02 °C@+200 °C
Circulator (SL-C)
SL
Temperature selection
digital
via
keypad
indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY (°C/°F)
Resolution
°C
0.01
ATC3
INT / EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2.0
Heater wattage (at 115 V)
kW
1.0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
Max. pressure
at 0 liters
bar
0.7
Max. suction
at 0 liters
bar
0.4
Ambient temperature
°C
5 ... 40
Mains power connection 230 V/50-60 Hz
V/ Hz
190 ... 253 / 50-60
Current draw
(at 208 V / 230 V)
A
12 / 13
Mains power connection 115 V/50-60 Hz
V/ Hz
90 ... 127 / 50-60
Current draw
(at 100 V / 115 V)
A
9/10
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C