Supply

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused

2)

.

Supply voltage

at the gauge

ripple


+13 … +32 VDC
Class 2 / LPS
≤50 mV

pp

Current consumption

<20 mA

Power consumption

≤0.3 W

Fuse required

2)

1 AT (slow), automatic reset
(Polyfuse)

The gauge is protected against reverse polarity of the supply
voltage.

Electrical connection

3CAx-xxx-0xxx
3CAx-xxx-1xxx
3CAx-xxx-Axxx


FCC68/RJ45, 8-pin, socket
D-Sub, 9-pin, male
Binder M12, 8-pin, male

Sensor cable

3CAx-xxx-0xxx
3CAx-xxx-1xxx
3CAx-xxx-Axxx


8-pin plus shielding
9-pin plus shielding
8-pin plus shielding

Cable length

≤100 m (0.14 mm² conductor)

For longer cables, larger conductor cross-sections are required.

Grounding concept

Vacuum flange - signal common

→ "Power Connection"

Supply common - signal common conducted separately; for dif-

ferential measurement (10

Ω)

2)