Assembly objects (class code 04 hex ), Output assemblies, Input assemblies – INFICON PEG100-D Medium to Ultra-High Vacuum Gauge User Manual
Page 6: 3 assembly objects (class code 04

6
tina12e1 (0106) PEG100-D.ga
Attribut ID
Access Rule
Name
Data/Type
Description
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
55 (37
hex
)
get
Serial Number
String [5]
56 (38
hex
)
get
Device Configuration
String [8]
PEG100-D
57 (38
hex
)
get
Device Status
UNIT
Device status
1 = Initialising
2 = Idle (HV on)
4 = Executing (HV on)
58 (3A
hex
)
get / set
Reporting Mode
BYTE
Polling, Bit Strobe = 6
COS / Cyclic = 0
60 (3C
hex
)
get
Exception Status
BYTE
0
hex
= ok
1
hex
= HV on, no plasma
2
hex
= HV off
Services
Service Code
Name
Description
–––––––––––––––––––––––––––––––––––––––––––––––––
14 (0E
hex
)
Get Attribute Single
3.3 Assembly Objects (Class Code 04
hex
)
A collection of assembly objects allows the sending of attributes from different application objects in one message
(i.e.: Polling I/O).
3.3.1 Output Assemblies
Messages which a master sends to the PEG100-D.
Output Assembly 1
Byte
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
0
res
res
res
res
res
res
HV
HV
ON / OFF
ON / OFF
Control
3.3.2 Input Assemblies
Messages which the PEG100-D sends to the master.
Input Assembly 2
Byte
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
0
res
res
res
res
res
HV
HV Status
Sensor
ON / OFF
Satus
Source
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1
Exception Status
2
Pressure value (Low Byte)
3
Pressure value (Low Middle Byte)
4
Pressure value (High Middle Byte)
5
Pressure value (High Byte)