Purpose, Technical data, General data – INFICON PEG100 Medium to Ultra-High Vacuum Gauge User Manual
Page 4: 2 technical data
1.1.1 Purpose
The PEG100 Penning Gauge is a compact active pres-
sure converter housing a measurement system as well
as the corresponding operating electronics. It has been
developed specifically for integration into vacuum
systems and offers a measurement range from 1·10
-9
to
1·10
-2
mbar.
The PEG100 is connected directly to the vacuum sys-
tem through its DN 25 KF or DN 40 CF flange.
The electrical connection is provided through a screened
8-way FCC 68 connector.
Moreover, the technical data as published in Section 1.2
must be observed.
1.2
Technical Data
1.2.1 General Data
Measurement range
1·10
-9
to 1·10
-2
mbar/Torr
Measurement uncertainty in the range
from 1·10
-8
to 1·10
-4
mbar:
Deviation from the characteristic
± 30 % of the displayed value
Average temperature coefficient of the output span
< 0.5 % / K of the displayed value
Reproducibility
< 4 % of the displayed value
Measurement principle
Cold cathode ionization according to Penning
Supply voltage
14.5 to 36 V DC, typ. 24 V DC
Ripple
≤
2 V
pp
Power consumption
< 2 W
Protection
IP 40
Electromagnetic compatibility (EMI)
CE mark
interference tolerance to EN 50082-2, Tab. 1, 2, 3
interference emission levels to EN 50081-1 and
FCC Rules Part 15, Class B
Flammability
UL 94 - V2
4
tina14e1 (0106) PEG100.ga
1.1.1 Purpose
The PEG100 Penning Gauge is a compact active pres-
sure converter housing a measurement system as well
as the corresponding operating electronics. It has been
developed specifically for integration into vacuum
systems and offers a measurement range from 1·10
-9
to
1·10
-2
mbar.
The PEG100 is connected directly to the vacuum sys-
tem through its DN 25 KF or DN 40 CF flange.
The electrical connection is provided through a screened
8-way FCC 68 connector.
Moreover, the technical data as published in Section 1.2
must be observed.
1.2
Technical Data
1.2.1 General Data
Measurement range
1·10
-9
to 1·10
-2
mbar/Torr
Measurement uncertainty in the range
from 1·10
-8
to 1·10
-4
mbar:
Deviation from the characteristic
± 30 % of the displayed value
Average temperature coefficient of the output span
< 0.5 % / K of the displayed value
Reproducibility
< 4 % of the displayed value
Measurement principle
Cold cathode ionization according to Penning
Supply voltage
14.5 to 36 V DC, typ. 24 V DC
Ripple
≤
2 V
pp
Power consumption
< 2 W
Protection
IP 40
Electromagnetic compatibility (EMI)
CE mark
interference tolerance to EN 50082-2, Tab. 1, 2, 3
interference emission levels to EN 50081-1 and
FCC Rules Part 15, Class B
Flammability
UL 94 - V2
4
tina14e1 (0106) PEG100.ga