INFICON SKY CDG160D 1 ... 1000 Torr / mbar User Manual
Page 27
tina53e1-c (2011-11)
27
It also permits to adjust an offset of the characteristic curve in
order to
• compensate for the offset of the measuring system or
• obtain a slightly positive zero for a 0 … 10 V AD converter.
The offset should not exceed 2% of the F.S. (+200 mV). At a
higher positive offset, the upper limit of the measurement range
is exceeded.
Zero adjustment using the ramp function can be
performed via
• the
• the diagnostic port (→ [4]),
• the RS232C interface (→ [4]).
Recommended procedure for adjusting the offset of a
measuring system:
→ Notice 24.
Operate the gauge for at least 2 hours under constant
ambient conditions (until the signal is stable).
Push the
keep it depressed. The
After 5 s, the zero adjustment value, starting at the current
output value, keeps continually changing (ramp) until the
button is released or until the setting limit (max. 25% F.S.)
is reached. The corresponding output signal is delayed by
about 1 s.