N removing and replacing the uv lamp, Age 84, Battery pack care – INFICON 2020ComboPRO Portable Photoionization Detector User Manual
Page 91
84
ISP-074-575-P1D
84
Battery Pack Care
Leaving the 2020ComboPRO for more than 3 months without charging the batteries
may result in the loss of recorded data and setup parameters. If the 2020ComboPRO
will not be used for long periods of time, recharge the battery for a few hours every
30 days to avoid loss of data. See Battery Charging on page 39 for instructions on
charging the battery.
Please observe the following:
For replacement battery pack use only Part No. MX700010 or MX700011.
Do not dispose of the battery pack in a fire. The cell may explode.
The MX700010 battery pack is 24% cadmium (Cd) by weight. This battery
pack must be disposed of properly. Check with local codes for special disposal
instructions.
The MX700011 battery pack is NIMH – Nickel Metal Hydride. This pack does
not contain cadmium. This battery pack must be disposed of properly. Check
with local codes for special disposal instructions.
Do not open, repair, or mutilate the battery pack.
Exercise care in handling battery packs in order not to short the terminals with
conducting materials such as rings, bracelets and keys. The battery or conductor
may overheat and cause burns.
Charge the battery pack using the AC adapter provided with or identified for use
with this product only in accordance with the instructions and limitations
specified in this manual. For AC adapter use only Part No. MX350002 (North
America) or MX396013 (Europe).
Maintenance of the UV Lamp
Removing and Replacing the UV Lamp
NOTE: Do not remove or replace the UV lamp in a hazardous location.
To remove the UV lamp from your instrument:
1. Ensure the instrument is turned off.
CAUTION
You must turn off the instrument before removing the lamp cover.