The 2020combopro intrinsic safety (i/s) notice – INFICON 2020ComboPRO Portable Photoionization Detector User Manual
Page 22
![background image](/manuals/562486/22/background.png)
ISP-074-575-P1D
15
15
The 2020ComboPRO Intrinsic Safety (I/S) Notice
THE 2020ComboPRO IS CLASSIFIED FOR USE IN CLASS I, DIVISION 1,
GROUPS A, B, C, D HAZARDOUS LOCATIONS. T4 (135°C) RATING.
It has been listed by MET® to comply with Underwriters Laboratories® Inc. UL®
913 Standard for Intrinsically Safe Apparatus and Associated Apparatus for use in
Class I, Division 1, Groups A, B, C, D Hazardous (Classified) Locations, Sixth
Edition when powered by MX700010 / MX700011 Battery Pack. THE
2020ComboPRO IS NOT INTENDED TO DETECT COMBUSTIBLE LEVELS
OF GASES. THE 2020ComboPRO
IS CLASSIFIED FOR USE IN
ATMOSPHERES CONTAINING COMBUSTIBLE LEVELS OF GASES.
Dette tilbehør må anvendes med 2020ComboPRO i EX- klassificeret område i
henhold til ovenfor beskrevet:
MX350006
Kalibreringsgas Regulator
MX350007
Håndrem
MX700260
Bæltetaske
MX700260
Field kit taske.
M1260147
Bruger reference kort
MX395001
Kort sampleprobe
Anvend ikke andet tilbehør til 2020ComboPRO i EX-klassificeret område.
Anvendes ikke originale komponenter kan det påvirke
sikkerhedsklassificeringen.
BEMÆRK
For minimering af risiko for brand samt skade på personer
læs og følg denne instruktion:
1. Kalibrering, vedligeholdelse, service, batteriskift og
opladning af batterier må ikke foretages i EX-
klassificeret område.
2. Det er kun batteripakke, UV lampe og sample filter,
som skiftes af burger.
3. Der er indvendig i 2020ComboPRO ikke udskiftelige
servicedele for bruger.