beautypg.com

Transport and storage, Troubleshooting – Heidolph EKT Hei-Con G User Manual

Page 26

background image

26

E

TRANSPORT AND STORAGE

Prior to transport

Switch off the magnetic stirrer.

Remove the vessel from the hotplate.

Pull the 5-pole DIN plug (K) out of the socket on the back of the magnetic stirrer.

Remove the EKT Hei-Con and the measuring sensor from the holders.

Push the holders off the fixation rod.

Unscrew the fixation rod (C) from the housing of the magnetic stirrer.

Screw the black cap (E) onto the top of the magnetic stirrer again.

Transport and storage

1.

The instrument and its components can be stored best in their original packaging or

other suitable container to avoid any damage during transport. Close the packaging with
adhesive tape.

2.

Store the instrument in a dry place.

TROUBLESHOOTING

No heating function:

EKT Hei-Con is not plugged in

Display “Hi“ or “Lo“ > security shutdown of the heating due to faulty sensor

Display “Max ERROR“ > security shutdown of the heating due to surpassing of
security shutdown temperature “Max” (25K over set temperature) Examine the EKT
(during servicing)

Maximum temperature limiter for the heating plate of the magnetic stirrer was

activated
(allows heating plate to cool down)

EKT Hei-Con is faulty (broken sensor)

Maximum-temperature limiter of the hotplate has actuated

(allow hotplate to cool down)

Caution
When transporting the instrument, please avoid any knocks, jolts
and vibrations

This manual is related to the following products: