Disposal

(Please note current regulations)

Parts of the rotary vane pumps and compressors,

shafts, iron or aluminum castings, plastic or glass

parts or bearings, may be recycled as scrap

materials.

PUMP DISASSEMBLY

1. Remove the 6 cap screws from the pump end

plate and remove the end plate.
DO NOT at any time remove the rotor or loosen
any of the electric motor thru-bolts.

2. Check for scoring on End Plate, Rotor, and

Body. Surfaces should be smooth. If severe

scoring is visible contact an Authorized Service

Facility.

3. Remove vanes.

4. Clean all surfaces with Gast recommended

Flushing solvent and dry well.

PUMP RE-ASSEMBLY

5. Apply a light coat of Gast recommended Oil

part# AD220 (an equivalent 10 wt. high

detergent oil can also be used) to the vanes.

6. Re-install vanes, noting the proper direction of

the beveled edge (Refer to exploded view).

7. Install End Plate with the six (6) cap screws.

Torque the screws to 100 lb.-in.

If the pump fails to produce proper vacuum or pressure,

or is excessively noisy, turn off and return unit to an

Authorized Service Facility.

Gast will NOT guarantee field-rebuilt product

performance. For performance guarantee, the product

must be returned to a Gast Authorized Service Facility.
Service Kit contents vary. Most contain vanes, gaskets

oiler wick and filter elements.

SERVICE KIT INSTALLATION

Disconnect electrical power supply cord before
installing Service Kit.

If product is hard wired into system, disconnect
electrical power at the circuit breaker or fuse box
before installing Service Kit.

Disconnect air supply and vent all air lines to
release pressure or vacuum.

Failure to follow these instructions can result in
death, fire or electrical shock.

WARNING

Electrical Shock Hazard

Do Not attempt to remove the rotor. It is held in place

by Loctite and can only be serviced by a Gast

Authorized Service Facility.

We have Gast Certified Service
Centers throughout the world. For
the most up-to-date listing, contact
one of our sales offices below:

World Headquarters
P.O. Box 97

2300 Highway M139

Benton Harbor, MI 49022

Ph: 269/926-6171

FAX: 269/925-8288

www.gastmfg.com

Gast Group Limited, United Kingdom
Unit 11, The I O Centre

Nash Road

Redditch, B98 7AS

United Kingdom

ph: +44 (0) 1527 504040

Fax: +44 (0) 1527 525262

www.gastmfg.com

Gast Hong Kong
Unit 12, 21/F, Block B

New Trade Plaza

6, On Ping Street, Shatin

N. T. Hong Kong

Ph: (852) 2690 1066

Fax: (852) 2690 1012

www.gasthk.com

This manual is related to the following products: