Relief Valve (hidden)

Turn the regulator screw

clockwise to increase pressure.

Pressure Gauge

The pressure gauge shows the

output pressure..

10 Micron Filter Element

Check muffler element each time oil

reservoir is filled; clean with soap and

water as needed, or replace. (See page 7

for Ordering Information).

GAST LUBRICATED LABORATORY 23 SERIES ROTARY

VANE VACUUM PUMPS & COMPRESSORS

O

PERATION

& M

AINTENANCE

M

ANUAL

www.gastmfg.com

ISO 9001 & 14001 CERTIFIED

®Registered Trademark/™Trademark of Gast Manufacturing Inc., Copyright © 2003 Gast Manufacturing Inc. All Rights Reserved.

• Operate at 32ºF - 104ºF (0ºC - 40ºC).

• Protect unit from dirt & moisture.

• Protect all surrounding items from hot exhaust

air. This exhaust air can become very hot.

• Product is not a consumer product and is for

commercial applications only.

• Do not pump flammable or explosive gases or use

in an atmosphere that contains such gases.

• Corrosive gases and particulate material will

damage unit. Water vapor, oil-based contaminants

or other liquids must be filtered out.

• Consult your Gast Distributor/Representative

before using at high altitudes.

• Use Gast #AD220 or a detergent SAE#10

automotive engine oil for lubricating.

PART NO. 70-350 G386PL (Rev. F)

WARNING

PLEASE READ THIS MANUAL COMPLETELY BEFORE INSTALLING AND USING

THIS MOTOR. SAVE THIS MANUAL FOR FUTURE REFERENCE AND

KEEP IN THE VICINITY OF THE MOTOR.

Product Use Criteria:

Relief Valve

Turn the regulator screw clockwise to

increase vacuum. NOTE: Maximum

pressure and vacuum cannot be

produced at the same time.

Vacuum Gauge

The vacuum gauge shows the intake

pressure.

Siphon Oiler

Models 0323, 0523: Check the oil level at

each use to ensure safe operation.

Constant Level Oiler (not shown)

Models 0823, 1023: Large oil reservoir;

check the oil level at longer intervals.

10 Micron Filter Element

Check filter element each time oil

reservoir is filled; clean with soap and

water as needed, or replace. (See

page 7 for Ordering Information).

NOTE: Removes solid particles only.

Install a moisture trap if liquids could

be drawn into the pump.