Gast DOA UL60601-1 User Manual
Oil-less diaphragm, Warning, Peration
OIL-LESS DIAPHRAGM
VACUUM PUMPS & COMPRESSORS
RECOGNIZED TO UL60601-1 REQUIREMENTS
FOR MEDICAL AND DENTAL APPLICATIONS
O
PERATION
& M
AINTENANCE
M
ANUAL
Product Use, Criteria, and Purpose:
• Pump can be used as compressor or vacuum.
• Max permissible working pressure is limited to 60 PSI.
• Max vacuum 26 IN.-Hg.
• Pump is class I type (B) equipment.
• Pump not suitable for use in the presence of a flammable
anesthetic mixture with air, oxygen or nitrous oxide.
• Pump is provided with hospital grade cord and plug.
• Pump only clean, dry air.
• Protect unit from dirt & moisture.
• This pump is oil-less and requires NO lubrication.
• Degree of protection against the ingress of water
(IPXO).
• Pump was not tested for (EMC) Elecro-magnetic
compatability.
Thank you for purchasing this Gast product. It is manufactured to the highest standards
using quality materials. Please follow all recommended maintenance, operational
and safety instructions and you will receive years of trouble free service.
www.gastmfg.com
ISO 9001 & 14001 CERTIFIED
®Registered Trademark/™Trademark of Gast Manufacturing Inc., Copyright © 2002 Gast Manufacturing Inc. All Rights Reserved.
Model DOL (DOA) For Medical & Dental
1
PLEASE READ THIS MANUAL COMPLETELY BEFORE INSTALLING AND USING
THIS PRODUCT. SAVE THIS MANUAL FOR FUTURE REFERENCE AND
KEEP IN THE VICINITY OF THE PRODUCT
WARNING
PART NO. 70-3006 G444PL (Rev. B)
2300 Highway M139 • Benton Harbor, MI 49022 • 269-926-6171
Electrical Specifications:
• External fast acting fuse, quantity 2 (3.15 AMP) for
115VAC. DOL (DOA)-MXXXX-AA rated amp 4.2.
• External fast acting fuse, quantity 2 (2.0 AMP) for
230VAC. DOL (DOA)-MXXXX-BN rated amp 2.2.
Transportation/Storage Conditions:
• Temperature range -28.5oC / -20oF to 50oC / 122oF.
• Relative humidity 10% to 90% (non-condensing).
• Barometric pressure minimum of 49K Pascal.
Packaging:
Unit is set in 3 cardboard die-cuts to fit sides and bottom
of unit. Unit is then saran wrapped and placed in
cardboard box. Two cardboard spacers are put on
each side after placed in box and then box is sealed
shut.