View calibration values, Key-in zero calibration value – DIGIWEIGH DWP-10000F Series User Manual
Page 17
![background image](/manuals/557890/17/background.png)
16
8. Repeat 2-4 using different test weight value and End C3 then revert to H17.
9. If the calibration was successful, revert to H17.
10. If the calibration was not successful, one of the error messages below will appear.
Take the indicated action to correct the problem then perform a new calibration.
"Err0" -The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"
Err1
" -
The calibration test weight or the adjusted keyed-in weight is smaller than 1% of
the full capacity of the scale. Change the calibration test weight or check the input data.
"Err2" -The internal resolution of the scale is not high enough to accept the calibration
value.
View calibration values
Note: The values displayed in this procedure are valid only after a successful calibration
has been performed using H16 and H17.
1. While in the Setup mode, scroll to "H 18", then scroll down once using the ZERO
key to enter view calibration menu.
2. The display will momentarily show "CAL 0" followed by a value. This value is the
zero calibration value and should be recorded in the table below. Press any key to
continue.
3. The display will momentarily show "CAL 1" followed by another value. This value is
the span calibration value and should also be recorded in the table below. if H15=2
or H15=3,it can also show “CAL 2” or “CAL 3”. Press any key to return to upper level
(H18).
Key-in zero calibration value
Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid zero calibration value, obtained from a successful H16 calibration
procedure, must be used.
1. While in the Setup mode, scroll to "H 19" then scroll down once using the ZERO
key.
2. The display will momentarily show "CAL 0", followed by a flashing zero. Use the
four directional keys (shown in Figure 6-1) to adjust the displayed value to the zero
calibration value.
3. After setting the exact value, press the NET/GROSS key to save the value.
4. The display will show "E CAL 0" momentarily then revert back up to H19.