beautypg.com

Det-Tronics PIRVOLVR Volumetric Infrared Hydrocarbon Gas Process Monitor for Vapor Recovery Systems User Manual

Page 11

background image

9

95-8522

7 .1

DISaSSEMBLy aND CLEaNINg pROCEDURE
The PIRVOLVR should be inspected periodically to ensure

that its performance is not impaired by fouled optics or by

clogging of the sample draw cup or hydrophobic filter.

Inspection and/or periodic maintenance involves three

different areas.

IMPORTANT

Remove power before disconnecting and
removing the PIRVOLVR for maintenance.

Sample Draw Cup.

Perform a visual inspection of the

sample draw cup and associated plumbing, and clean

as required. External filtration of sample stream is

recommended.

hydrophobic Filter.

The flow of gas through the

filter can be inhibited by an accumulation of mud or

other contaminants. To inspect the hydrophobic filter,

disassemble the PIRVOLVR as described below. If the

filter appears to be fouled, replace it.

Optics.

Cleaning of the optical surfaces is required

only if an optical fault is indicated (1.0 mA current output

signal from the PIRVOLVR). This procedure is most easily

accomplished on a bench.

Required materials: Clean, flat work surface, foam tipped

swabs (no cotton), isopropyl alcohol, screwdriver.

CAUTION

The PIRVOLVR contains semiconductor
devices that are susceptible to damage by
electrostatic discharge. An electrostatic charge
can build up on the skin and discharge when
an object is touched. Therefore, use caution
when handling the device, taking care not to
touch electronic components or terminals. If
the electronics assembly is removed, it should
be placed in an anti-static bag or box while
stored or transported. A static safeguarded
work area is highly recommended (if available)
for disassembly and cleaning of the PIRVOLVR.

1. Remove the sample draw cup from the front of the

PIRVOLVR by pulling it straight off the electronics

assembly (no tools needed). See Figure 12.

2. Perform a visual inspection of the sample draw

cup and associated plumbing, and clean as

required. Also inspect the two O-rings inside the

sample draw cup and clean, lubricate or replace

as required.

3. Loosen the two captive screws on top of the

mirror assembly (Figure 12) and remove the

mirror assembly and hydrophobic filter from the

electronics assembly.

4. Thoroughly douse the interior of the mirror

assembly as well as the foam tipped swab

with isopropyl alcohol. Use the swab to gently

cleanse the surfaces of the reflecting mirrors

inside the mirror assembly. After cleaning with

the swab, flush out the mirror assembly using

a liberal amount of isopropyl alcohol. Tip the

mirror assembly with mirror openings downward

to remove accumulated isopropyl alcohol and

particle contaminants. Repeat the alcohol flush

to remove any remaining contaminants. Allow the

mirror assembly to air dry in a dust-free location.

IMPORTANT

Do not insert any sharp object into the mirror
assembly. Scratching of the mirrors will void the
PIRVOLVR warranty. Do not use cotton tipped
swabs or buds as they are likely to leave fiber
residue.

5. Clean the two windows on the electronics

assembly using the procedure described above.

When the mirror assembly and electronics

assembly are thoroughly dry, proceed with

re-assembly.

6. Slide a new hydrophobic filter over the end of the

mirror assembly, then carefully seat the mirror

assembly and hydrophobic filter against the base

of the electronics assembly. Be careful not to

crumple or fold the hydrophobic filter.

7. Tighten the two captive screws on the top of the

mirror assembly. Tighten the screws evenly. Do

not over-tighten (6 inch-pounds recommended).

8. Slide the sample draw cup back onto the

PIRVOL-VR.

9. Check for leaks around the PIRVOLVR sample

draw cup.