RAD Oven Series Owner

’s Manual

22

Version 1

Copyright © 2012 by Despatch Industries.

All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.

Notice


Review oven weight to ensure the foundation is adequate to hold
the weight.

4.2.1.1. Placement Requirements

Place oven directly on the floor.

Plumb and level the oven to assure proper heat distribution and operation of all mechanical

components.

Do not expose oven to excessive vibration and affix all electrical cabinets.

If placing the oven in an area where excessive particulate matter exists (such as a construction

site or a coal processing center), periodically clean all its electrical compartments. Keep
power supply within the specifications provided by Despatch, and use a line conditioner for a
facility with an unstable power supply.


4.3. Exhaust Connections

The RAD oven exhaust port is located on the top of the oven (Figure 6). Install an exhaust stack
from the discharge opening to the outside of the building. The discharge opening is 17.8 cm (7.0
inches) diameter.

Warning!


Do not use the oven in wet, corrosive or explosive atmospheres
unless this oven is specifically designed for a special
atmosphere.


All stacks must comply with state and local building codes to
insure that surrounding combustible surfaces are below 71°C
(160°F).

Warning!


Do not be careless about restrictions in and around the fresh air
and exhaust openings and stacks.

Notice


Flashing through roof or wall must be capable of handling
temperatures up to 343°C (650°F).