Temperature scale conversion (c/f) – Despatch LAC2-12-5 User Manual

Page 80

background image

71

APPENDIX: Temperature Scale
Conversion and Optional MRC5000
Setup

Temperature Scale Conversion (C/F)


The Protocol Plus controller can be operated in either C or F. The default setting for the
controller is C. Changing from one to the other is as follows:

1. Go into the Setup Mode on the controller.

2. Press the Select Key until Setup is displayed.

3. Press the Page key and Security will be displayed.

4. Press the Menu Key and Password will be displayed. Use the arrow keys to

enter the proper password. The default password is 2 for level two.

5. Once the proper password is displayed, press the Page key until PID is

displayed.

6. Press the Menu key and Temp Unit along with C or F will be displayed. Use the

arrow keys to change the setting.

7. Once the proper setting is displayed, press and hold the Page key for

approximately three seconds to exit the Setup Mode.