Table 8 Difficulties, Probable Causes, and Remedies
Difficulty
Probable cause
Suggested Remedy
Failure to heat
CONTROL instrument
malfunction
Check parameter in controller or replace
controller if OUT LED on controller does not turn
on.
Erratic
temperature
CONTROL instrument
malfunction
Check control parameters before replacing the
CONTROL instrument.
Inaccurate
temperature
CONTROL instrument
miscalibration
Recalibrate CONTROL instrument.
Oven will not
control at
setpoint
HI-LIMIT instrument set
too low
Set the HI-LIMIT higher.
CONTROL instrument
malfunction
Check control parameters before replacing
CONTROL instrument.
Air friction of recirculation
fan
Open exhaust air vent. Unit will not control at
minimum operating temperature with vent(s)
closed.
Heater does
not shutdown
until temp.
reaches the
HI-LIMIT
setting
CONTROL instrument
malfunction
Verify control parameters. Replace relay if no
output exists. Replace CONTROL instrument if
5VDC output exists.
SSR Relay malfunction
Replace relay if no CONTROL output exists.