Set-up parameters – Despatch DES2000 Controller User Manual
Page 12
6
Set-Up Parameters
1. Go into the operating parameters and change the LOC parameter to
–1 and
press the SET/ENT key to access the Setup Parameters.
Table 3 Set-Up Parameters for Degrees Centigrade
Table 4 Set-Up Parameters for Degrees Fahrenheit
Table 5 Oven Models and Temperatures
Oven Model
Degrees F
Degrees C
Maximum
Minimum
Maximum
Minimum
LAC
500
95
260
35
LAD/LFD/LND/LXD
650
95
343
35
LBB
400
86
204
30
NOTE: If the
“In” code is changed, the “P” code in the operating parameters may have
to be reset.
2. Press and hold the
key for three seconds to return to normal operation.
Code
Parameter Name
Setting
In
Input Type
5
SPH
Maximum Setpoint Limit
Use value from Table 5 (below)
SPL
Minimum Setpoint Limit
Use value from Table 5 (below)
SC
Super Function
OFF
dr
Direct/Reverse Action
0
DSP
Display Priority
0
Code
Parameter Name
Setting
In
Input Type
35
SPH
Maximum Setpoint Limit
Use value from Table 5 (below)
SPL
Minimum Setpoint Limit
Use value from Table 5 (below)
SC
Super Function
OFF
dr
Direct/Reverse Action
0
DSP
Display Priority
0