6

Set-Up Parameters

1. Go into the operating parameters and change the LOC parameter to

–1 and

press the SET/ENT key to access the Setup Parameters.

Table 3 Set-Up Parameters for Degrees Centigrade












Table 4 Set-Up Parameters for Degrees Fahrenheit










Table 5 Oven Models and Temperatures

Oven Model

Degrees F

Degrees C

Maximum

Minimum

Maximum

Minimum

LAC

500

95

260

35

LAD/LFD/LND/LXD

650

95

343

35

LBB

400

86

204

30



NOTE: If the

“In” code is changed, the “P” code in the operating parameters may have

to be reset.

2. Press and hold the

key for three seconds to return to normal operation.

Code

Parameter Name

Setting

In

Input Type

5

SPH

Maximum Setpoint Limit

Use value from Table 5 (below)

SPL

Minimum Setpoint Limit

Use value from Table 5 (below)

SC

Super Function

OFF

dr

Direct/Reverse Action

0

DSP

Display Priority

0

Code

Parameter Name

Setting

In

Input Type

35

SPH

Maximum Setpoint Limit

Use value from Table 5 (below)

SPL

Minimum Setpoint Limit

Use value from Table 5 (below)

SC

Super Function

OFF

dr

Direct/Reverse Action

0

DSP

Display Priority

0