Changing display from °c to °f – Despatch DES2000 Controller User Manual
Page 14
8
Changing Display From °C To °F
The control can be configured for either °C or °F. Use the following steps to change
control from displaying °C to °F.
1. Press and hold the
for three (3) seconds.
2. The display will read CtL. The SP LED will flash
indicating that Operating Parameter mode has been
entered.
3. Press the
until LoC appears on the display.
4. Press the
▲ or ▼once to enter the parameter.
5. Press the
▼ to set the value to –1. The SP LED will flash rapidly, indicating that
the Set-Up Parameter mode has been entered.
6. Press the
to enter the value. The display will read In.
7. Press the
▲or ▼ once to enter the parameter.
8. Press the
▲ to set the value to 35. The right decimal point LED will flash
indicating that the setting is being changed. This will stop flashing when the new
value has been entered.
9. Press the
to enter the value.
10. Press the
until SPH is displayed.
11. Press the
▲ to set the value to 400.
12. Enter a value of 400.
13. Press the
to enter the value.
14. Repeat steps 12
–14 for SPL; the value is 32.
15. Press and hold the
for three (3) seconds to return to the operation mode.
WARNING: Make sure
you understand what
you are changing
before doing so.
Changing the program
parameters will alter the
functions of the
CONTROL.
NOTE: To change from
°F to °C repeat the
above steps. The
following are the values
to be entered:
In = 5
SPH = 204
SPL = 0
P = 5