beautypg.com

Control Company 6412 PLATINUM ULTRA-ACCURATE THERMOMETER User Manual

Control Company Thermometers

background image

RECALL MINIMUM/MAXIMUM

1. Press MIN/MAX to view the minimum and maximum

temperatures recorded since turning the unit on or

since clearing min/max.

2. Press the MIN/MAX button once to display the

minimum temperature. The lower portion of the

display shows “MIN” indicating that this is the

minimum reading. While in the MIN/MAX mode,

select the desired format with the °F/°C or

.0/.00/.000/.0000 buttons.

3. Press the MIN/MAX button again to display the

Maximum temperature. The lower portion of the

display shows “MAX” indicating that this is the

maximum reading. A third press will return the

display to the current reading.

4. MIN/MAX may be reviewed at any time.

5. To clear the MIN/MAX, place the unit in normal mode

(not reviewing the MIN/MAX) and press MIN/MAX

and CLEAR simultaneously. You may also clear

MIN/MAX by turning the unit off.

HOLD FUNCTION

1. Press the HOLD button once to “freeze” the display

at the current temperature reading. “HOLD” appears

on the lower portion of the display indicating that the

unit is in hold mode.

2. While in the HOLD mode, select the desired format

with the °F/°C or .0/.00/.000 buttons.

3. Press the HOLD button a second time to return to

the current temperature reading. “HOLD” will no

longer appear on the display.

POWER

Do not turn the unit on and off rapidly. It may lock the

display. When turning the unit on/off the microprocessor

may become locked, inoperable, or the display may

read “888888.” If this occurs, reset the thermometer by

removing the battery, waiting 15 seconds, and replacing

the battery.

OPERATION

1. Plug the probe into the receptacle located on the right

side of the unit.

2. Press the ON/OFF button to turn the unit on.

3. Press the °F/°C button to change the display

between Fahrenheit and Celsius.

4. Press the .0/.00/.000/.0000 button to change the

display to read the desired resolution.

5. Use the probe to monitor temperatures in air/gas,

liquids, and semi-solids. Place the stainless-steel

portion of probe in contact with the material to be

measured. In most instances the depth of the probe

needs to be approximately ½ inch.

6. Press the ON/OFF button to turn the unit off when the

thermometer is not in use to prolong battery life.

HISTORY MODE

History provides an effortless method to observe when a

temperature change takes place. It continuously displays

for the past 24 hours starting with 1 hour ago. If left on for

more than 24 hours, it displays only the most recent 24

hours. History may be reviewed at any time.

1. After 1 hour, press the HISTORY button once to

show the current or first hour reading. The display

will show “1” on the far right to indicate this is the first

reading. After approximately two seconds, the display

will show the temperature and “min” which indicates

this is the minimum reading for the first hour. After

approximately two more seconds, the display will

again show “1”. The display continues to alternate

between these two displays. While in the history

mode, the °F/°C or .0/.00/.000 buttons may be used

to select the desired format.

2. The second press of the HISTORY button shows the

maximum temperature for the current or first hour.

After approximately two more seconds, the display

will show “1”.

3. With each press of the HISTORY button, the unit

will scroll through all 48 minimum and maximum

readings.

4. To exit the history mode, press the CLEAR button

or MIN/MAX button. As long as the unit is in the

history mode, pressing the CLEAR button alone will

not clear the history.

5. To clear history, first exit the history mode and then

press HISTORY and CLEAR buttons simultaneously.

Turning the unit off does not clear the history.

SPECIFICATIONS

Range:

-328.00 to 932.000°F

-200.00 to 500.000°C

Resolution:

0.001°, 0.01° or 0.1°

(user selectable)

from -99 to 932°F/-99 to 500°C

0.01° or 0.1°

from -328 to -100°F/-200 to -100°C

0.0001° from -9.9999 to 99.9999 °F/°C

Sampling Rate: approximately 2 times per second

Memories:

48 Memories are stored in the History

Mode (min/max for the past 24 hours)

2 Memories are stored in the Min/Max

Mode (min/max for the current hour)

Probes Supplied:

Model 6412 supplied with a fast response, handle-probe.

The stainless steel portion of the probe is capable of

measuring temperatures through the entire range (-328 to

932°F/-200 to 500°C)

Model 6413 supplied with a fast response, bullet-probe

with PTFE cable. The sensor and cable are capable of

being subjected to temperatures of -148 to 392°F/-100

to 200°C

BUTTON DESCRIPTION

ON/OFF

Turns unit on and off.

°F/°C:

Selects °Fahrenheit/°Celsius

.0/.00/.000/.0000 Selects decimal point on display

HISTORY

Displays history memories

HOLD/CLEAR Press to hold reading also used in

conjunction with HISTORY or

MIN/MAX buttons to exit the mode or

clear readings.

MIN/MAX

Displays the highest and lowest

temperatures recorded for the current

hour.

DISPLAY MESSAGES

▲ Appearing indicates temperature is rising.

▼ Appearing indicates temperature is falling.

“LLL” Displayed when the temperature being

measured is below the range of the unit or when

there is an open probe or no probe.

“HHH” Displayed when the temperature being

measured is above the range of the unit or

when there is a shorted probe.

“BAT” Displayed when the battery is low and needs to

be replaced. See Battery section for battery

replacement instructions.

RECEPTACLES

The receptacle on the right side of the unit is for the

probe. The upper left side receptacle is for an accessory

AC Adaptor. The lower left side receptacle is for the

accessory Data Acquisition System data cable.

DATA OUTPUT

In order to use the data output, the Accessory Data

Cable Cat No. 4099 is required. See Accessories section

for ordering information.

With the accessory data cable attached, to toggle the

RS232 output on/off, press and hold the °F/°C and

.0/.00/.000/.0000 simultaneously (2320 indicates

RS232 output is “on”, 232F indicates RS232 output

is “off”).

BATTERY

If “BAT” appears on the display, it indicates the

batteries are low and need to be replaced. To replace

the battery, slide the battery cover located on the back

of the unit away from the unit. Remove the old battery

and replace it with a new 9-Volt alkaline battery. Use an

alkaline battery, NOT a regular or heavy duty battery.

Properly connect the battery. Replace the battery cover.

Incorrectly installed batteries may damage electronics.

ALL OPERATIONAL DIFFICULTIES

If this thermometer does not function property for any

reason please replace the battery with a new 9-Volt

alkaline battery Battery section, above). Low battery

power can occasionally cause any number of “apparent”

operational difficulties. Replacing the battery with a new

fresh battery will solve most difficulties.