Control Company 4052 LONG-STEM THERMOMETER User Manual
Traceable® long-stem thermometer instructions
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail [email protected]
www.control3.com
Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025 accredited as a Calibration Laboratory by A2LA.
Cat. No. 4052 / 4352/ 4353 / 4354
Traceable® Digital Long-Stem Thermometer Instructions
92-4052-10
Rev. 3 040809
Traceable® is a registered trademark of Control Company
©
2009
Control Company
TRACEABLE® LONG-STEM THERMOMETER INSTRUCTIONS
Display:
3/8” LCD
Resolution:
0.1° above –20 and below 200
1.0° below –20 and above 200
Sampling Rate: 10 seconds in Fahrenheit mode
1 second in Celsius mode
Features:
Stainless steel probe
protective plastic sleeve
OPERATION
1. To turn the thermometer on slide the ON/OFF
switch to the ON position.
2. To measure in Fahrenheit select the “F” position
of the F/C switch or “C” to measure in Celsius.
3. Remove the protective plastic sleeve of the
thermometer probe and insert the probe into the
material to be measured.
4. To turn the thermometer off slide the ON/OFF
switch to the OFF position.
To conserve battery life, always turn the unit off
when not in use.
SPECIFICATIONS:
Model:
4052
4352
4353
4354
Range:
–58 to 302°F
–58 to 302°F
–58 to 572°F
–58 to 572°F
–50 to150°C
–50 to 150°C
–50 to 300°C
–50 to 300°C
Accuracy:
±1°C
±0.2°C
±1°C
±0.5°C
Length
Stem:
8”
8”
11.4”
11.4”
Overall:
11”
11”
14.4”
14.4”
Diameter:
Stem:
1/8”
1/8”
1/8”
1/8”
Top:
1.2”
1.2”
1.2”
1.2”
ALL OPERATIONAL DIFFICULTIES
If this unit does not function properly for any rea-
son, replace the battery with a new high-quality
battery (see “Battery Replacement” section). Low
battery power can occasionally cause any number
of “apparent” operational difficulties. Replacing the
battery with a new fresh battery will solve most dif-
ficulties.
BATTERY REPLACEMENT
Erratic readings, a faint display, or no display, are
all indicators that the battery must be replaced.
Remove the battery cap, located at the top of
the unit by turning it in the direction of the arrow
(counterclockwise). Remove the exhausted battery
and replace with a new 1.5 volt battery. Replace the
battery cover. Do not overtighten the cap. Replace-
ment battery Cat. No. 1039.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact: