beautypg.com

Caron 6241 User Manual

Page 27

background image

6200 series Operations Manual

Rev D 2/14/2012

Page 27 of 60

Analog Outputs:

With DLUX301 & DLUX302 controllers, there is an analog output signal for

temperature and humidity which represents the actual chamber values. This allows

the chamber to be connected to an in-house data acquisition or alarm system.

The analog signal outputs are selectable as either voltage DC or milliamp. In both

cases, the output is scalable from 0.0 to 20.0. Common settings are 0-1V, 0-5V, 1-5V,

0-10V, 0-20mA, and 4-20mA. The factory default settings are 0-5V.

For each selected output range, a temperature and humidity value must correspond to

the high and low range. This range should be large enough to encompass the entire

chamber possible values and small enough to provide adequate resolution. The

analog outputs can be calibrated by placing an offset into the controller. This offset

affects only the analog outputs and not the controller displayed value. To change the

controller displayed value, see the calibration section in the manual.

To change the factory defaults of the controllers, the controllers must be unlocked.

Refer to Section 14, Unlocking the Controllers.

1. Pressing the up and down keys simultaneously for 6 seconds.

2. Press the up key until otPt is displayed

3. Press the advance key until the upper display reads 1

4. Press the up button until 3 is displayed

5. Press the advance button to scroll through the parameters; use the up and down

arrow buttons to change the parameters. Factory default parameters are listed

below

Parameter Description

Display

Value

Type

o.tY

Volt

Function

Fn

rMt

Retransmit Source

r.Sr

Ai

Scale Low

S.Lo

0.00

Scale High

S.hi

5.00

Range Low

r.Lo

0.0

Range High

r.hi

100.0

Calibration Offset

o.CA

0.0

For a DC volt output, set the Type to “Volt”

For a mA output, set the Type to “MA”

This manual is related to the following products: