6200 series Operations Manual

Rev D 2/14/2012

Page 19 of 60

Incoming compressed air must be free of dirt, oil,

moisture and filtered to 10 microns minimum.

Incoming compressed air must not exceed 100

psig.

CARON recommends installing a pressure gauge, filter, and shutoff prior to the dryer

to monitor incoming air pressure. The air supply (either house air or other compressed

air source) must be 90 to 100psig at the air supply inlet. An oil-less compressor may

be used. A Sound Suppression Kit maybe used to reduce the dryer noise (contact

CARON service for details).

1. Attach Dryer Package to exterior chamber with four mounting screws.

2. Connect dryer package power cord and 3/8” orange tubing into the dryer panel

on the chamber

Mounting

Screws

Mounting

Screws

Power Cord

Connection

Tube

Connection