Caron 6241 User Manual
Page 22
6200 series Operations Manual
Rev D 2/14/2012
Page 22 of 60
Installing Drain Water Pump (PUMP301)
In applications where a floor drain is not available and a CARON water recycling
system is not being used, a drain pump can be purchased to pump any excess
condensate from the chamber to a local sink or drain. The pump is located near the
middle of the back of the chamber. Connect the supplied tubing from the pump to the
sink / drain. The tubing may be run vertically into a ceiling but should not exceed 15
feet height.
Installing Side Mounted Recorders (RCDR314 & RCDR315)
The recorder will arrive packaged inside the chamber. Carefully remove the recorder
from its packaging. Mount the recorder by using the pre-installed recorder bracket.
There are three factory drilled holes located on the right side of chamber as you face
the front of chamber. Using the factory supplied screws, screw the recorder to the side
of the chamber.
There are two cables that come out of the recorder. One is to power the recorder; the
other is the temperature and/or humidity signals coming from the chamber. With
power to the equipment turned off, plug the two connectors into their mating connector
at the top of the chamber. Turn power to the chamber back on.
Standard factory set up for chart speed is 7 day operation. Refer to the Chart
Recorder’s User’s Manual provided with the recorder to change the chart speed
settings for various chart speeds.