beautypg.com

Caron 6241 User Manual

Page 13

background image

6200 series Operations Manual

Rev D 2/14/2012

Page 13 of 60

Connect each chamber to a grounded circuit.

Failure to do so could result in electrical shock.

Use only distilled or deionized water with a

resistivity between 50K

-CM and 1M

-CM and

a pH of greater than 6.5. Using water outside

this range will void your warranty.

Do not use water that contains chloramines.

Chloramines can damage internal rubber

gaskets resulting in leaks.

A water inlet fitting on the back of the unit and ¼” black tubing are provided to connect

the water supply to the chamber. Connect an appropriate water supply to the fitting.

Incoming line pressure should be regulated to not exceed 80 psi.

If a Condensate Recirculator water recycling system was purchased as a water supply,

refer to its user’s manual for proper installation of the water supply.

Connecting Electrical Power

The unit requires a dedicated electrical outlet. See table below for model specific

power required and connection.

Model #

Power Requirements

Plug Connection

-2

230V, 60Hz, 12A FLA

NEMA 6-15

-3

230V, 50Hz, 10A FLA

CEE 7/7

When the required electrical connection is available, plug the provided power cord into

the unit and the electrical outlet.

This manual is related to the following products: