Operating
16
LMC20-GB
Delay timer with ramp rate functions
Timer Ended
When ramp rate active
Indicator
– RUN = On - if ramp rate active
Scrolling Text
– Timer Running
To HOLD timer
Press and quickly
release
+
Indicator
– RUN = Flashing
Scrolling Display
– Timer Hold
Static Text -
OFF
To switch off
Timer / Cancel
Press and Hold
+
for more than 1
second
Indicator
– RUN = On - if ramp rate active
Scrolling Display
– None
Static Text -
OFF
To re-RUN Timer
Press and quickly
release
+
Indicator
– RUN = On
Scrolling Display
– Timer Running
Static Text
– OFF
Power Failure While Using Delay Timer
If there is a power failure while the timer is running and the power is subsequently restored
the timer will reset and will re-run from the power on time.
6.3
Soft start Timer
Operation
Action
Indication
To RUN Timer
Press and quickly
release
+
Indicator
– RUN = On
Scrolling Display
– Timer Running
Static Text -
OFF
To HOLD timer
Press and quickly
release
+
Indicator
– RUN = Flashing
Scrolling Display
– Timer Hold
Static Text -
OFF
To switch off
Timer / Cancel
Press and Hold
+
for more than 1
second
Indicator
– RUN = Off
Scrolling Display
– None
Static Text -
OFF
Timer Ended
Indicator
– RUN = Off
Scrolling Text
– Timer End
Running Soft Start Timer
When the soft start timer is set to run, the scrolling display will indicate “timer running” and
the set point temperature, not the soft start set point.
Power Failure While Using Soft Start Timer
If there is a power failure while the timer is running and the power is subsequently restored
the timer will reset and will re-run from the power on time.
Ramp Rate with Soft Start Timer
It is not recommended that the ramp rate function be used with a soft start timer.
Note: If the temperature is already above the threshold when the timer is set to run the timer
will time out immediately.