Operating

10

LMC20-GB

4.3

Customer ID.


A furnace or oven identification number can be entered if required. This maybe used to
identify one of many units, for production or quality control systems.

Press scroll

 until the display shows ID <customer ID>. Using the Up 

Down

 enter

your own identification number. This can range from 1-9999.

4.4

Units

Press scroll

 until the display shows units <Display Units>. Using the Up 

or Down

select the required units.

Mnemonic

Description

nonE

No units (Default °C)

°C

Celsius

°F

Fahrenheit

°K

Kelvin

PErc

%

4.5

Language

The scrolling text on the 3216 can be shown in different languages, this can only be set at
the factory and therefore must be specified at the time of placing an order.

5 Timer

5.1

Setting the Timer


A timer can be configured to operate in four different modes. These can be selected in level
2 (supervisor level) using the TM.CFG parameter as:-

None

Dwell Timer

Delayed switch on timer

Soft start timer


None

The timer is turned off, no timer configurations are available, the instrument works as a
simple set point controller.

Press scroll

 until the display shows TM.CFG <TIMER CONFIGURATION>. Using the Up 

Down

 select NONE.