Fuses & power control – Carbolite BLF 18/21 User Manual
Page 15
BLF
MF23 – 3.11
15
F
USES
&
P
OWER
C
ONTROL
8.5 Fuses
8.6 F1 to F3 - refer to the circuit diagrams
F1
Internal supply
fuses
Fitted if supply cable fitted.
GEC Safeclip of the type shown
(glass type F up to 16A)
F2
Auxiliary
circuit fuses
2 Amps glass type F
32mm x 6mm
F3
Heat Light
fuses
2 Amps glass type F
32mm x 6mm
8.7 Thyristor
Fuse
Ferraz Protistor of the rating
shown
Customer
fuses
Required if no supply cable fitted.
Recommended if cable fitted.
See rating label for amperage; see
table below for fuse rating.
The table below gives the fuse rating for the various models.
8.8 Power Control
The setting for the power limit parameter in the controller (
OP.Hi
) should be 100% for all models.
The thyristor stacks are set to the secondary (element) maximum currents given in the table.
TABLE OF FUSE AND CURRENT LIMIT VALUES
Model
Phases
volts
supply fuse
rating
type
thyristor
fuse
rating
current
limit
(element
circuit)
BLF 17/3
1-phase
220-240V
30-32A NS32
50A
90A
BLF 18/3
1-phase
220-240V
30-32A NS32
50A
150A
BLF 17/8
3-phase+N
380/220-415/240V 16A
glass
25A
82A
BLF 17/8
3-phase delta
208, 220-240V
30-32A NS32
25A
82A
BLF 18/8
3-phase+N
380/220-415/240V 16A
glass
25A
150A
BLF 18/8
3-phase delta
208, 220-240V
30-32A NS32
25A
150A
BLF 17/21
BLF 18/21