beautypg.com

Fuses & power control – Carbolite BLF 18/21 User Manual

Page 15

background image


BLF


MF23 – 3.11

15

F

USES

&

P

OWER

C

ONTROL

8.5 Fuses

8.6 F1 to F3 - refer to the circuit diagrams

F1

Internal supply
fuses

Fitted if supply cable fitted.

GEC Safeclip of the type shown
(glass type F up to 16A)

F2

Auxiliary
circuit fuses


2 Amps glass type F
32mm x 6mm

F3

Heat Light
fuses

2 Amps glass type F
32mm x 6mm

8.7 Thyristor

Fuse

Ferraz Protistor of the rating
shown

Customer
fuses

Required if no supply cable fitted.
Recommended if cable fitted.

See rating label for amperage; see
table below for fuse rating.

The table below gives the fuse rating for the various models.

8.8 Power Control

The setting for the power limit parameter in the controller (

OP.Hi

) should be 100% for all models.

The thyristor stacks are set to the secondary (element) maximum currents given in the table.

TABLE OF FUSE AND CURRENT LIMIT VALUES

Model

Phases

volts

supply fuse
rating

type

thyristor
fuse
rating

current
limit
(element
circuit)

BLF 17/3

1-phase

220-240V

30-32A NS32

50A

90A

BLF 18/3

1-phase

220-240V

30-32A NS32

50A

150A

BLF 17/8

3-phase+N

380/220-415/240V 16A

glass

25A

82A

BLF 17/8

3-phase delta

208, 220-240V

30-32A NS32

25A

82A

BLF 18/8

3-phase+N

380/220-415/240V 16A

glass

25A

150A

BLF 18/8

3-phase delta

208, 220-240V

30-32A NS32

25A

150A

BLF 17/21

BLF 18/21

This manual is related to the following products: