Fuses & power control – Carbolite HTF Series User Manual
Page 14
RHF & HTF
14
MF22 3.09
8.0 F
USES
&
P
OWER
C
ONTROL
8.1 Fuses
F1-F2: Refer to the circuit diagrams.
F1
Internal supply
fuses
Fitted if supply cable fitted.
GEC Safeclip of the type shown
(glass type F up to 16A)
F2
Auxiliary
circuit fuses
2 Amps glass type F
32mm x 6mm
F3
Heat Light
fuses
2 Amps glass type F
32mm x 6mm
Thyristor Fuse
Ferraz Protistor of the rating
shown
Customer
fuses
Required if no supply cable fitted.
Recommended if cable fitted.
See rating label for amperage; see
table below for fuse rating.
The table below gives the fuse rating for the various models.
8.2 Power Control
The setting for the power limit parameter in the controller (
OP.Hi
) should be 100% for all models.
The thyristor stacks are set to the secondary (element) maximum currents given in the table.
TABLE OF FUSE AND CURRENT LIMIT VALUES
Model
Phases
volts
supply fuse
rating
type
thyristor
fuse
rating
current
limit
(element
circuit)
RHF 17/25
3-phase + N
380/220-415/240
25A
NS25
30A
150A
RHF 17/25
3-phase delta
208, 220-240
40A
ES40
30A
150A
HTF 18/15
1-phase
220-240
50A
ES50
50A
150A
HTF 18/15
3-phase + N
380/220-415/240
15-16A type F1
20A
150A
HTF 18/15
3-phase delta
220-240
30-32A NS32
50A
150A
HTF 18/27
3-phase + N
380/220-415/240
30-32A NS32
50A
160A
HTF 18/27
3-phase delta
220-240
63A
ES63
50A
160A