7.5 Fuses

F1-F3: Refer to the circuit diagrams.

F1

Internal
supply
fuses

Fitted if supply cable fitted.
Fitted on board to some types of EMC filter.

on-board and up to 16 Amps:
32mm x 6mm type F
other: GEC Safeclip

F2

Auxiliary
circuit
fuses

Fitted on board to some types of EMC filter.
May be omitted up to 25Amp/phase supply
rating.

2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm

F3

Heat Light
fuses

May be omitted up to 25 Amp/phase supply
rating.

2 Amps glass type F
32mm x 6mm

Customer
fuses

Required if no supply cable fitted.
Recommended if cable fitted.

See rating label for amperage;
see table below for fuse rating.

Model

phases

Volts

Supply Fuse
Rating

Volts

Supply Fuse
Rating (type)

CFM 12/1, 12/2

1-phase

220-240

5A

110-120

10A

CFM 14/1, 14/2

1-phase

220-240

15A or 16A

110-120

30A or 32A

CFM 14/1 + Aux

1-phase

220-240

16A

CFM 14/2 + Aux

1-phase

220-240

20A

CFM 14/4

1-phase

220-240

20A

110-120

40A

Note: For technical reasons any internal fuses fitted may have a higher raing than the
recommended supply fuses.

7.6 Power Settings

The power limit settings (parameter

OP.Hi

) for the 1200°C CFM models are all 100%, with the

exception of any units supplied for 254V operation where the limit is 89%.

The power limit settings for the 1400°C models are as follows. The figures represent the maximum

percentage of time that controlled power is supplied to the elements. Do not attempt to “improve

performance” by setting a value higher than the correct one from the table.

Model

Volts:

110V 120V

208V

220V

230V

240V

254V

CFM 14/1

67

56

89

67

61

56

50

CFM 14/2

67

56

89

67

61

56

50

CFM 14/4

100

100

92

82

Auxiliary heater models: the power limits for the controllers for the auxiliary heaters have power
limits of 100%..