C.B.S. œ Scientific

Double & Triple Wide Mini-Vertical

6

1.3

Safety

Power to the Double and Triple Wide Mini-Vertical Gel units is to be supplied by an external DC
voltage power supply that must be ground isolated so that the DC voltage output floats with
respect to ground. For any power supply used, the maximum specified operating parameters for
the units are:

Maximum Limits

Single Unit

Dual

Unit

250 VDC

250 VDC (10cm), 700 VDC (16cm)

20

watts

power

40

watts

power

75mA

current

150mA

current

50

°C ambient temperature

50

°C ambient temperature

Current to the unit, provided from the external power supply, must enter the unit through the lid

assembly, providing safety interlock to the user. Current to the unit is broken when the lid is
removed. Do not attempt to use the unit without the safety lid, and always turn the power
supply off before removing the lid, or when working with the unit in any way. Follow
safety precautions specified by the power supply manufacturer.