beautypg.com

2 electro- blotting procedure, 3 removing the blot – C.B.S. Scientific QNX-700 User Manual

Page 16

background image

16

Quadra Instructions 9-30-13

4. Insert blotting casettes into core making sure that

red side faces outward. See diagram 5.

5. Close doors and re-latch by pressing down on

the white latches so that assembly looks like that
shown in figure 6. If running one blot, slide white
reservoir conversion plate into the side without
the blotting cassette.

5.2

Electro- Blotting Procedure

1. Place stirring bar in bottom corral of lower res-

ervoir. Place core/blotting cassette assemblies

into lower reservoir. The anode (red) and cath-

ode (black) electrodes are color-coded on both

the core/cassette assembly and lower reservoir.

Ensure the red dot on the cassette assemblies

are on the same side as the red receptacle on the
lower reservoir.

Attach safety cover and turn on

magnetic stirrer. If you have the cooled unit, cat.
#QNC-700, precool buffers and attach external
cooler (set to the desired temperature) to the
tubing adaptors so that the inlet is on the bottom
and the outlet is on top (see fig. 9 on page 10).

2. Pour 1.7 liters of freshly prepared, chilled (4º) buf-

fer into lower buffer reservoir. Buffer will percolate

into central core.

3. Attach safety cover. The unit should look as

shown in figure 7 and is ready for power.

4. Connect the leads to the power supply, matching

the color-coded red to red and black to black as

shown in figure 8.

See Section 6.2 for recom-

mended power conditions. Begin transfer by
electrophoresis.

5.3

Removing the Blot

1.

Turn the power supply off and disconnect
the leads from the power supply.
Remove the
safety cover from the unit, by placing thumbs on
white posts next to red & black connectors, then
pushing down while pulling up with fingers under
lid

. Do not remove safety cover by pulling up

on leads!

2. Blotting cassettes can be removed by leaving the

core in place and opening the top latches of the
core, opening the doors and lifting the cassettes
out. Unlatch the blotting cassettes and remove
blot from blotting sandwich.

5

red side

black sides

6

8

7

This manual is related to the following products: