OUT OF RANGE SPAN
Out of Range Span The Out of Range Span was designed to eliminate extrane-
ous signals and noise. This Span or window changes with the distance of the
average distance reading. This window will allow only readings that fall within
its limits. If the target is suddenly changed outside of the window limits, the
sensor will wait until a number or echoes specified in the Out of Range Count
are received within a new window before updating the output reading.
To Read the Out of Range Span in a sensor, use the command structure:
Send
Receive
AAh
Start Byte
66h
Start Byte
01h,
Sensor Number
01h
Sensor Number
11h
Command
11h
Command
00h
Data (not used)
##h
Span (High Byte)
00h
Data (not used)
##h
Span (Low Byte)
55h
End Byte
##h
NA
99h
Stop Byte
To change the Out of Range Span in the sensor to 304 mm, use the command
structure;
Send
AAh
Start Byte
01h,
Sensor Number
12h
Command
01h
Data
304 decimal = 130h
30h
Data
55h
End Byte