beautypg.com

APG PWS Paddle Wheel user manual User Manual

Page 12

background image

PWS Series

Rev. A1, 4/07

12

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

• Shaft and spindle is damaged.

Remedies

• Connect the power.
• Repair power supply
• Wire correctly
• Calibrate sensitivity or replace paddle
• Install sensor in good location
• Replace unit.

Symptom

Paddle is not covered by materials. Relay energized.

Possible Causes

• Insuffi

cient power supplied.

• Miswiring
• Heavy deposit on shaft
• Material has dead stock
• Temperature too high
• Shaft is bent or damaged.

Remedies

• Connect the power.
• Repair power supply
• Wire correctly
• Calibrate sensitivity or replace paddle
• Install sensor in good location
• Replace unit.

Symptom

Relay chatters

Possible Causes

• Insuffi

cient power supplied

• Loose cables
• Material too fl uid

Remedies

• Repair power supply
• Tighten connections
• Use a time delay relay.