9
enter into the sterilization procedure. The Default time is as following:
1. Modo 1
Vacuum: 10 min
Ozone generating: 30 min
Ozone-maintain: 10 min
Ozone decomposition: 20 min
The whole sterilization time is 70 min.
2. Modo 2
Vacuum: 10 min
Ozone generating: 25 min
Ozone-maintain: 10 min
Ozone decomposition: 15 min
The whole sterilization time is 60 min.
NOTES:
After start, it shows TIME and SIGN of each stage by turns. The
procedure and sign of each stage is as following:
Vacuum (CC) Ozone generating (O3) Ozone-maintain (OF)
Ozone decomposition (O2) OVER (OP).
During the auto mode procedure, any press on thin film will be no
function.
It shows
“OP” when whole procedure is completed and buzz 3 times
at the same time. After, turn off the main switch.